Description
Calibration fixture for the L-700 Laser, L-703, T-212 4-Axis Target and T-261 4-Axis Target. Allows the calibration of the V & H center values and the V & H angular values using cal-lab-certified micrometers and HLI software. Can also be used to set the laser beam to the middle if it’s adjustment range prior to starting an alignment job. Used with the S-1390 Spindle8 and S-1410 R-1307 Calibration software programs.
2 models depending on the laser:
- A-807 – used with L-700, L-702/L-702SP and L-703 lasers
- A-807-705 – used with L-705 and L-706 lasers
Applications
- Spindle alignment of lathes, turning centers, grinders
- Calibration of most of HLI’s bore or spindle targets
Features
- .001 in (0.025 mm) calibrated micrometer with resolution up to .0001 in. (0.0025 mm) for measuring PSD center span errors.
- .0001 in. (0.025 mm) calibrated micrometer built-in tilt stage with 6 in. (152.4 mm) tilt plate for PSD measuring angular errors.
- Used to generate calibration factors for HLI’s alignment software.
- 2 models depending on the laser:
- A-807 – used with L-700, L-702/L-702SP and L-703 lasers
- A-807-705 – used with L-705 and L-706 lasers
Specifications
Dimensions:
L17 x H9.5 x W3 in. (431 x 241 x 76 mm)
Weight:
22.5 lbs oz (10.2 kg)
Distance between mounting surfaces:
12 in (302 mm)
Mounting Hole Diameter:
0.4995 in (12.687 mm)
Center Calibration Resolution:
.0001 in. (0.003 mm) with .001 in. calibrated micrometer
Angular Calibration Resolution:
.0001 in. per 6 in. (0.003 mm per 152.4 mm ) with .0001 in. calibrated micrometer